Recall Ledger

CHARLOTTE CLAW Screw recall (Wright Medical Technology, Inc.)

Medical deviceClass IIOngoing

Summary: Wright Medical Technology, Inc. is recalling CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System. Reason: Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.

What is being recalled

CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System

Why

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO. (states named: California, Colorado, Florida, Idaho, Michigan, New Jersey, Oklahoma, Pennsylvania, Tennessee, Texas, Wisconsin)

How to tell if yours is affected

UDI-DI: 00840420109705, Lot: 1826666, Distributed: 18-Aug-2025 through 09-Apr-2026

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3334-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 18, 2026
ReportedOctober 7, 2026
CompanyWright Medical Technology, Inc.
Company locationarlington, TN, United States
Initiated byVoluntary: Firm initiated
Quantity34

More recalls from Wright Medical Technology, Inc. · all medical device recalls · recalls from August 2026